==Bring main and secondary NavX carts into room.==
 
==Bring main and secondary NavX carts into room.==
 
+
===Secondary cart===
 
*Mount magnet to table plug side towards the foot of the table
 
*Mount magnet to table plug side towards the foot of the table
 
[[File:NavX1.jpg|thumb|150px|left|NavX magnet plug]]
 
[[File:NavX1.jpg|thumb|150px|left|NavX magnet plug]]
Line 11: Line 11:
 
*Plug power cord into power outlet
 
*Plug power cord into power outlet
 
[[File:NavX6.jpg|thumb|150px|left|Power]]<br style="clear:both" />
 
[[File:NavX6.jpg|thumb|150px|left|Power]]<br style="clear:both" />
*Ground and fiber optic cables
+
*Ground, fiber, and data cables
[[File:NavX8.jpg|thumb|150px|left|Ground and fiber cables]]<br style="clear:both" />
+
[[File:NavX8.jpg|thumb|150px|left|Ground, data, fiber cables]]<br style="clear:both" />
Ground cable plugs into grounding block
+
*Ground cable plugs into grounding block
 
[[File:NavX9.jpg|thumb|150px|left|Ground block]]<br style="clear:both" />
 
[[File:NavX9.jpg|thumb|150px|left|Ground block]]<br style="clear:both" />
 
*Two Fiber optic cables connect as shown. Orange cables on top, gray under
 
*Two Fiber optic cables connect as shown. Orange cables on top, gray under
 
[[File:NavX10.jpg|thumb|150px|left|Fiber optic location]]<br style="clear:both" />
 
[[File:NavX10.jpg|thumb|150px|left|Fiber optic location]]<br style="clear:both" />
 +
*Serial data cable connects to left of fiber optic cables
 +
[[File:NavX11.jpg|thumb|150px|left|Serial data cable]]<br style="clear:both" />
  
 
+
===Main Cart===
 
+
*From second drawer of main cart unpack splitter unit
 
+
[[File:NavX12.jpg|thumb|150px|left|Splitter unit]]<br style="clear:both" />
 +
*Splitter unit is installed between pin-blocks and amplifier, also between ecg cable and amplifier
 +
[[File:NavX13.jpg|thumb|150px|left|Splitter unit location]]<br style="clear:both" />
 +
*Unplug A,B,C cables from amplifier
 +
[[File:NavX14.jpg|thumb|150px|left|Unplug pin-blocks]]<br style="clear:both" />
 +
*Replace with A,B,C cables from splitter
 +
*Repeat with ecg cable
 +
*A,B,C and ecg cables plug into splitter
 +
[[File:NavX15.jpg|thumb|150px|left|A,B,C cables go here]]<br style="clear:both" />
 +
*Colored plugs from splitter connect to St. Jude amplifier on main cart
 +
[[File:NavX16.jpg|thumb|150px|left|Splitter cables]]<br style="clear:both" />
 +
*From top drawer of main cart remove patch connection unit
 +
[[File:NavX17.jpg|thumb|150px|left|Patch unit]]<br style="clear:both" />
 +
*Secure patch unit over table control as shown
 +
[[File:NavX18.jpg|thumb|150px|left|patch unit location]]<br style="clear:both" />
 +
*Patch unit plugs into amplifier here
 +
[[File:NavX19.jpg|thumb|150px|left|Patch unit plug location]]<br style="clear:both" />
 +
*Locate power, ground, and fiber cables on side of main cart
 +
[[File:NavX20.jpg|thumb|150px|left|Cables on side of main cart]]<br style="clear:both" />
 +
*Plug in power cord
 +
[[File:NavX21.jpg|thumb|150px|left|Power cord]]<br style="clear:both" />
 
*Clip ground to other ground cables or suitable metal object
 
*Clip ground to other ground cables or suitable metal object
 
[[File:NavX7.jpg|thumb|150px|left|Ground]]<br style="clear:both" />
 
[[File:NavX7.jpg|thumb|150px|left|Ground]]<br style="clear:both" />
 
+
*One end of fiber optic cable is plugged into bottom fiber port below plugs from earlier step marked "Precision"
 
+
[[File:NavX22.jpg|thumb|150px|left|Fiber cable plug]]<br style="clear:both" />
 
+
*Other end of fiber cable plugs into St. Jude amplifier port marked "Precision"
 
+
[[File:NavX23.jpg|thumb|150px|left|Precision amplifier port]]<br style="clear:both" />
[[File:STCarto11.jpg|thumb|150px|left|Power and ground cables]]<br style="clear:both" />
+
===Secondary cart part 2===
 +
*Identify fiber cable on back of secondary cart
 +
[[File:NavX24.jpg|thumb|150px|left|Fiber cable secondary cart]]<br style="clear:both" />
 +
*Plug this cable into fiber port on St. Jude amplifier
 +
[[File:NavX25.jpg|thumb|150px|left|Fiber plug on amplifier]]<br style="clear:both" />
 +
*Identify black Ethernet cable near Maquette amplifier marked "Abbott Ensite"
 +
*Plug into back of Quartz unit on secondary cart
 +
[[File:NavX26.jpg|thumb|150px|left|Quartz unit Ethernet plug]]<br style="clear:both" />
 +
*From front of ablator on secondary cart, identify quad cable
 +
[[File:NavX27.jpg|thumb|150px|left|Ablator quad cable]]<br style="clear:both" />
 +
*Plug pins into A block 1-4
 +
[[File:NavX28.jpg|thumb|150px|left|Quad pins in C block]]<br style="clear:both" />
 +
*Identify large gray cable on back of secondary cart
 +
[[File:NavX29.jpg|thumb|150px|left|Large secondary cart cable]]<br style="clear:both" />
 +
*Plug into port on St. Jude amplifier
 +
[[File:NavX30.jpg|thumb|150px|left|St. Jude amplifier]]<br style="clear:both" />

Latest revision as of 17:19, 4 December 2023

Bring main and secondary NavX carts into room.

Secondary cart

  • Mount magnet to table plug side towards the foot of the table
NavX magnet plug
Mounted on table


  • Locate magnet cable on secondary cart and plug into magnet (arrows on plug will be toward the floor)
Magnet cable


Cable plug location


  • Locate power, ground, and fiber optic cable hanging on secondary NavX cart
Power, ground, fiber


  • Plug power cord into power outlet
Power


  • Ground, fiber, and data cables
Ground, data, fiber cables


  • Ground cable plugs into grounding block
Ground block


  • Two Fiber optic cables connect as shown. Orange cables on top, gray under
Fiber optic location


  • Serial data cable connects to left of fiber optic cables
Serial data cable


Main Cart

  • From second drawer of main cart unpack splitter unit
Splitter unit


  • Splitter unit is installed between pin-blocks and amplifier, also between ecg cable and amplifier
Splitter unit location


  • Unplug A,B,C cables from amplifier
Unplug pin-blocks


  • Replace with A,B,C cables from splitter
  • Repeat with ecg cable
  • A,B,C and ecg cables plug into splitter
A,B,C cables go here


  • Colored plugs from splitter connect to St. Jude amplifier on main cart
Splitter cables


  • From top drawer of main cart remove patch connection unit
Patch unit


  • Secure patch unit over table control as shown
patch unit location


  • Patch unit plugs into amplifier here
Patch unit plug location


  • Locate power, ground, and fiber cables on side of main cart
Cables on side of main cart


  • Plug in power cord
Power cord


  • Clip ground to other ground cables or suitable metal object
Ground


  • One end of fiber optic cable is plugged into bottom fiber port below plugs from earlier step marked "Precision"
Fiber cable plug


  • Other end of fiber cable plugs into St. Jude amplifier port marked "Precision"
Precision amplifier port


Secondary cart part 2

  • Identify fiber cable on back of secondary cart
Fiber cable secondary cart


  • Plug this cable into fiber port on St. Jude amplifier
Fiber plug on amplifier


  • Identify black Ethernet cable near Maquette amplifier marked "Abbott Ensite"
  • Plug into back of Quartz unit on secondary cart
Quartz unit Ethernet plug


  • From front of ablator on secondary cart, identify quad cable
Ablator quad cable


  • Plug pins into A block 1-4
Quad pins in C block


  • Identify large gray cable on back of secondary cart
Large secondary cart cable


  • Plug into port on St. Jude amplifier
St. Jude amplifier