Line 31: | Line 31: | ||
*A,B,C and ecg cables plug into splitter | *A,B,C and ecg cables plug into splitter | ||
[[File:NavX15.jpg|thumb|150px|left|A,B,C cables go here]]<br style="clear:both" /> | [[File:NavX15.jpg|thumb|150px|left|A,B,C cables go here]]<br style="clear:both" /> | ||
− | + | *Colored plugs from splitter connect to St. Jude amplifier on main cart | |
− | + | [[File:NavX16.jpg|thumb|150px|left|Splitter cables]]<br style="clear:both" /> | |
− | + | *From top drawer of main cart remove patch connection unit | |
− | + | [[File:NavX17.jpg|thumb|150px|left|Patch unit]]<br style="clear:both" /> | |
− | + | *Secure patch unit over table control as shown | |
+ | [[File:NavX18.jpg|thumb|150px|left|patch unit location]]<br style="clear:both" /> | ||
+ | *Patch unit plugs into amplifier here | ||
+ | [[File:NavX19.jpg|thumb|150px|left|Patch unit plug location]]<br style="clear:both" /> | ||
+ | *Locate power, ground, and fiber cables on side of main cart | ||
+ | [[File:NavX20.jpg|thumb|150px|left|Cables on side of main cart]]<br style="clear:both" /> | ||
+ | *Plug in power cord | ||
+ | [[File:NavX21.jpg|thumb|150px|left|Power cord]]<br style="clear:both" /> | ||
*Clip ground to other ground cables or suitable metal object | *Clip ground to other ground cables or suitable metal object | ||
[[File:NavX7.jpg|thumb|150px|left|Ground]]<br style="clear:both" /> | [[File:NavX7.jpg|thumb|150px|left|Ground]]<br style="clear:both" /> | ||
− | + | *One end of fiber optic cable is plugged into bottom fiber port below plugs from earlier step marked "Precision" | |
− | + | [[File:NavX22.jpg|thumb|150px|left|Fiber cable plug]]<br style="clear:both" /> | |
− | + | *Other end of fiber cable plugs into St. Jude amplifier port marked "Precision" | |
− | + | [[File:NavX23.jpg|thumb|150px|left|Precision amplifier port]]<br style="clear:both" /> | |
− | [[File: | + | ===Secondary cart part 2=== |
+ | *Identify fiber cable on back of secondary cart | ||
+ | [[File:NavX24.jpg|thumb|150px|left|Fiber cable secondary cart]]<br style="clear:both" /> | ||
+ | *Plug this cable into fiber port on St. Jude amplifier | ||
+ | [[File:NavX25.jpg|thumb|150px|left|Fiber plug on amplifier]]<br style="clear:both" /> | ||
+ | *Identify black Ethernet cable near Maquette amplifier marked "Abbott Ensite" | ||
+ | *Plug into back of Quartz unit on secondary cart | ||
+ | [[File:NavX26.jpg|thumb|150px|left|Quartz unit Ethernet plug]]<br style="clear:both" /> | ||
+ | *From front of ablator on secondary cart, identify quad cable | ||
+ | [[File:NavX27.jpg|thumb|150px|left|Ablator quad cable]]<br style="clear:both" /> | ||
+ | *Plug pins into A block 1-4 | ||
+ | [[File:NavX28.jpg|thumb|150px|left|Quad pins in C block]]<br style="clear:both" /> | ||
+ | *Identify large gray cable on back of secondary cart | ||
+ | [[File:NavX29.jpg|thumb|150px|left|Large secondary cart cable]]<br style="clear:both" /> | ||
+ | *Plug into port on St. Jude amplifier | ||
+ | [[File:NavX30.jpg|thumb|150px|left|St. Jude amplifier]]<br style="clear:both" /> |
Latest revision as of 17:19, 4 December 2023
Secondary cart
- Mount magnet to table plug side towards the foot of the table
- Locate magnet cable on secondary cart and plug into magnet (arrows on plug will be toward the floor)
- Locate power, ground, and fiber optic cable hanging on secondary NavX cart
- Plug power cord into power outlet
- Ground, fiber, and data cables
- Ground cable plugs into grounding block
- Two Fiber optic cables connect as shown. Orange cables on top, gray under
- Serial data cable connects to left of fiber optic cables
Main Cart
- From second drawer of main cart unpack splitter unit
- Splitter unit is installed between pin-blocks and amplifier, also between ecg cable and amplifier
- Unplug A,B,C cables from amplifier
- Replace with A,B,C cables from splitter
- Repeat with ecg cable
- A,B,C and ecg cables plug into splitter
- Colored plugs from splitter connect to St. Jude amplifier on main cart
- From top drawer of main cart remove patch connection unit
- Secure patch unit over table control as shown
- Patch unit plugs into amplifier here
- Locate power, ground, and fiber cables on side of main cart
- Plug in power cord
- Clip ground to other ground cables or suitable metal object
- One end of fiber optic cable is plugged into bottom fiber port below plugs from earlier step marked "Precision"
- Other end of fiber cable plugs into St. Jude amplifier port marked "Precision"
Secondary cart part 2
- Identify fiber cable on back of secondary cart
- Plug this cable into fiber port on St. Jude amplifier
- Identify black Ethernet cable near Maquette amplifier marked "Abbott Ensite"
- Plug into back of Quartz unit on secondary cart
- From front of ablator on secondary cart, identify quad cable
- Plug pins into A block 1-4
- Identify large gray cable on back of secondary cart
- Plug into port on St. Jude amplifier