Secondary cart
- Mount magnet to table plug side towards the foot of the table
- Locate magnet cable on secondary cart and plug into magnet (arrows on plug will be toward the floor)
- Locate power, ground, and fiber optic cable hanging on secondary NavX cart
- Plug power cord into power outlet
- Ground, fiber, and data cables
- Ground cable plugs into grounding block
- Two Fiber optic cables connect as shown. Orange cables on top, gray under
- Serial data cable connects to left of fiber optic cables
Main Cart
- From second drawer of main cart unpack splitter unit
- Splitter unit is installed between pin-blocks and amplifier, also between ecg cable and amplifier
- Unplug A,B,C cables from amplifier
- Replace with A,B,C cables from splitter
- Repeat with ecg cable
- A,B,C and ecg cables plug into splitter
- Colored plugs from splitter connect to St. Jude amplifier on main cart
- From top drawer of main cart remove patch connection unit
- Secure patch unit over table control as shown
- Patch unit plugs into amplifier here
- Clip ground to other ground cables or suitable metal object