==Procedure== | ==Procedure== | ||
− | === | + | ===Pre=== |
Read [[UCD EP Role]] for pre and post procedural steps. | Read [[UCD EP Role]] for pre and post procedural steps. | ||
*Include CMAP when patching for phrenic pacing | *Include CMAP when patching for phrenic pacing | ||
− | ** | + | **Cable located in C block #31-32 |
**Attach alligator clips to patches | **Attach alligator clips to patches | ||
*Set up pressure lines | *Set up pressure lines | ||
Line 125: | Line 125: | ||
**Vascular US | **Vascular US | ||
**EP cables | **EP cables | ||
+ | ***Achieve to pin block Blue 1-10 | ||
+ | ***Deca nav to Deca | ||
+ | ***His Quad to Quad B (OK to unplug pin block quad plug) | ||
+ | |||
===During case=== | ===During case=== | ||
*Open ablation window on live screen | *Open ablation window on live screen |
Revision as of 16:07, 11 October 2023
Open
- EP Pack
- Flexcath
- 28mm Arctic Advance
- Probe Cover
- 2 - Ice Cover - Shockwave
- Copilot
- Manifold + dump bag
- Smart Ablate Tubing
- Micropuncture
- 20ml syringes 2-4
- A-line set
- Pressure transducer kit x3
- 7F locking
- 8F locking
- 8.5 Non-locking
- 6fr CRD 2 quad
Pezeshkian | Srivatsa | Cortez | Munir |
---|---|---|---|
Stockroom | |||
Versacross D0 | Versacross D1 | ||
Soundstar 10Fr | Soundstar 8Fr | ||
DecaNav D | DecaNav F | ||
Lab | |||
Octaray D 2-2-2 | Octaray F 3-3-3 | ||
Dynamic XT steerable quad | CRD 2 Quad | ||
Achieve 20mm | Achieve 15mm | ||
Long 11Fr | Long 9Fr | ||
14fr non-locking | Zero Gravity drape | ||
Standby | If needed for touch up post cryo or for redo afib. | ||
Smart touch thermocool STSF DF | Smart touch thermocool ST FJ | ||
Perclose x1
Vascade x3 |
Cables
- Achieve cable
- Electrical umbilical cable
- Coaxial umbilical cable
- DecNav cable
- Pentaray cable
- Quad cable x 2
- Abalator cable
Procedure
Pre
Read UCD EP Role for pre and post procedural steps.
- Include CMAP when patching for phrenic pacing
- Cable located in C block #31-32
- Attach alligator clips to patches
- Set up pressure lines
- Double spike Hep saline bag under pressure
- Flush and connect blue pressure to transducer closest to patient on rack
- Zero A-line
- Assist with hooking up
- Vascular US
- EP cables
- Achieve to pin block Blue 1-10
- Deca nav to Deca
- His Quad to Quad B (OK to unplug pin block quad plug)
During case
- Open ablation window on live screen
- Check that ablation switch behind ablation consoles is set to "Cryo"
- Physician will place catheters and access LA.
- Measure LA pressure
- Measure conduction intervals
- Once Cryo ballon is in LA and ablation starts do the following:
- Save and label freeze frame of fluoro with location of freeze and ablation number ex: "LUPV Cryo #1" copy this text and save
- Monitor ablation sequence
- After ablation ends, record ablation data including copied text to Ablation End log entry comment section ex: "LUPV Cryo #1 180sec -45C"
- If esophagus temp is affected, record this temp also "23C esoph"
- EP study is conducted during Cryo ablation phase of procedure
- Atrial Wenckebach, Atrial ERP, and VERP
- When phrenic nerve is involved. Prepare to pace CMAP
- Channel 4 800msec, 25mA
- When freeze is initiated, come on pacing as temp passes 0C
- When thawing, come off pacing as temp passes 0C