UCD Cryo A-fib
From TheInterWiki
Open
- EP Pack
- Flexcath
- 28mm Arctic Advance
- Probe Cover
- 2 - Ice Cover - Shockwave
- Copilot
- Manifold + dump bag
- Smart Ablate Tubing
- Micropuncture
- 20ml syringes 2-4
- A-line set
- Pressure transducer kit x3
- 7F locking
- 8F locking
- 8.5 Non-locking
- 6fr CRD 2 quad
Pezeshkian | Srivatsa | Cortez | Munir |
---|---|---|---|
Stockroom | |||
Versacross D0 | Versacross D1 | ||
Soundstar 10Fr | Soundstar 8Fr | ||
DecaNav D | DecaNav F | ||
Lab | |||
Octaray D 2-2-2 | Octaray F 3-3-3 | ||
Dynamic XT steerable quad | CRD 2 Quad | ||
Achieve 20mm | Achieve 15mm | ||
Long 11Fr | Long 9Fr | ||
14fr non-locking | Zero Gravity drape | ||
Standby | If needed for touch up post cryo or for redo afib. | ||
Smart touch thermocool STSF DF | Smart touch thermocool ST FJ | ||
Perclose x1
Vascade x3 |
Cables
- Achieve cable
- Electrical umbilical cable
- Coaxial umbilical cable
- DecNav cable
- Pentaray cable
- Quad cable x 2
- Abalator cable
Procedure
Read UCD EP Role for pre and post procedural steps.
- Open ablation window on live screen
- Check that ablation switch behind ablation consoles is set to "Cryo"
- Physician will place catheters and access LA.
- Measure conduction intervals
- Once Cryo ballon is in LA and ablation starts do the following:
- Save and label freeze frame of fluoro with location of freeze and ablation number ex: "LUPV Cryo #1" copy this text and save
- Monitor ablation sequence
- After ablation ends, record ablation data including copied text to Ablation End log entry comment section ex: "LUPV Cryo #1 180sec -45C"
- If esophagus temp is affected, record this temp also "23C esoph"
- EP study is conducted during Cryo ablation phase of procedure
- Atrial Wenckebach, Atrial ERP, and VERP