UCD Micra/Aveir

From TheInterWiki
Revision as of 16:28, 25 October 2023 by Admin (talk | contribs)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)

Open

  1. EP pack
  2. Probe cover
  3. Micropuncture
  4. 8F non locking sheath
  5. Two 3mL transducers for Micra, Four 3mL transducers for Aveir
    • No 3 way stop cocks for Aveir cases
  6. Amplatz super stiff 0.35, 180cm
  7. Dilator set
  8. Two 60mL syringes
  • 20cc syringe
  1. Contrast
  2. 0-silk suture
  3. Steristrips

Standby

  1. Perclose × 2


Aveir steps

  1. US guided access with micropuncture
  2. Sheath wire
  3. 8fr. Dilator
  4. Preclose x2
  5. Amplatz SS loaded through second Perclose
  6. Dilate with 12, 16, 22 dilators
  7. Attach drip A to introducer sheath and insert
  8. Attach B, C, D drip lines to delivery system
    • Adjust drips and check for air
  9. Attach 20cc syringe with straight contrast to side port of C drip line
  10. Deploy device
  11. Close with Preclosed sutures
  12. Close skin with Woggle using 0-Prolene
  13. Dress with gauze and Tegaderm