Stereotaxis Setup

Home Magnets

  1. Bring C-arm in so arrows on ceiling line up
  2. Button # 2: Bring Magnets in until Chime.
  3. Button # 3: Stow magnets back.

During case

Register

  1. Catheter needs to be in body.
  2. Line up X-ray with arrows on ceiling.
  3. Bring magnets in (Button #2).
  4. Click on "Flame" icon (bottom left of big screen).
  5. Click on Carto 3 registration. Window will pop up.
  6. Might need to adjust table height.
  7. Biosense Stereotaxis mode on X-ray control screen.
  8. Click on X-ray transfer button on screen:
  9. 6 green dots should pop up on X-ray image. If not, step on X-ray in stereotaxis mode.
  10. Move sheath-marker on screen to proper position.
  11. When table height, FD, sheath marker in correct position, should be able to accept.
  12. For RAO x-ray overlay: click on x-ray transfer button on the top window.
  13. For LAO x-ray overlay: click on x-ray transfer button on the bottom window.
  14. Registration is complete.